Measurement error during filling

Fault description

Error pattern

Cause

Rectification

4. Measured value re-
mains unchanged du-
ring filling

Level

time

0

l False echoes in the close

range too big or product echo
too small

l Strong foam or spout genera-

tion

l Max. adjustment not correct

l Remove false echoes in the

close range

l Check measurement situation:

Antenna must protrude out of
the socket, installations

l Remove contamination on the

antenna

l Minimize interfering installa-

tions in the close range by
changing the polarization di-
rection

l Create a new false signal sup-

pression

l Adapt max. adjustment

5. Measured value re-
mains in the bottom
section during filling

Level

time

0

l Echo from the tank bottom

larger than the product echo,
for example, with products with
ε

r

<

2.5 oil-based, solvents

l Application parameter "Me-

dium", check Vessel height and
Floor form, adapt if necessary

6. Measured value re-
mains momentarily un-
changed during filling
and then jumps to the
correct level

Level

time

0

l Turbulence on the product

surface, quick filling

l Check application parameters,

change if necessary, e.g. in
dosing vessel, reactor

7. Measured value
jumps towards 0 % du-
ring filling

Level

time

0

l Amplitude of a multiple echo

(vessel top - product surface)
is larger than the level echo

l Check parameter "Application",

especially vessel top, product
type, dished end, high dielect-
ric figure, adapt if necessary

l The level echo cannot be

distinguished from the false
echo at a false echo position
(jumps to multiple echo)

l Remove/reduce false echo:

minimize interfering installa-
tions by changing the polariza-
tion direction

l Chose a more suitable instal-

lation position

8. Measured value
jumps towards 100 %
during filling

Level

time

0

l Due to strong turbulence and

foam generation during filling,
the amplitude of the product
echo sinks. Measured value
jumps to the false echo

l Carry out false signal suppres-

sion

9. Measured value
jumps sporadically to
100 % during filling

Level

time

0

l Varying condensation or con-

tamination on the antenna

l Carry out a false signal sup-

pression or increase false sig-
nal suppression with
condensation/contamination in
the close range by editing