VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) Modbus and Levelmaster protocol Approval according to LPR radio standard User Manual

Page 65

Advertising
background image

Ambient conditions

Ambient, storage and transport temperature

-40 … +80 °C (-40 … +176 °F)

Process conditions

For the process conditions, please also note the specifications on the type label. The lowest value
always applies.

Seal

Antenna impedance co-
ne

Process temperature (measured on the process
fitting)

FKM (SHS FPM 70C3
GLT)

PTFE

-40 … +130 °C (-40 … +266 °F)

PTFE

4)

-40 … +200 °C (-40 … +392 °F)

PEEK

5)

-40 … +200 °C (-40 … +392 °F)

FFKM (Kalrez 6375)

PTFE

-20 … +130 °C (-4 … +266 °F)

PEEK

-20 … +250 °C (-4 … +482 °F)

FFKM (Kalrez 2035)

PTFE

-15 … +130 °C (5 … +266 °F)

PEEK

-15 … +210 °C (5 … +410 °F)

FFKM (Kalrez 6230)

PTFE

-15 … +130 °C (5 … +266 °F)

PEEK

-15 … +250 °C (5 … +482 °F)

Graphite

Ceramic

-200 … +450 °C (-328 … +842 °F)

Graphite (process fitting
Hastelloy C 22)

Ceramic

-196 … +400 °C (-321 … +752 °F)

Vessel pressure - horn antenna

-

Antenna impedance cone PTFE

-1 … 40 bar (-100 … 4000 kPa/-14.5 … 580 psig)

-

Antenna impedance cone PP

-1 … 3 bar (-100 … 300 kPa/-14.5 … 43.5 psig)

-

Antenna impedance cone PEEK

-1 … 100 bar (-100 … 10000 kPa/-14.5 … 1450 psig)

-

Antenna impedance cone ceramic

-1 … 160 bar (-100 … 16000 kPa/-14.5 … 2320 psig)

Vessel pressure - parabolic antenna

-1 … 6 bar (-100 … 6000 kPa/-14.5 … 870 psig)

Vessel pressure relating to the flange nom-
inal stage

see supplementary instructions manual "Flanges
according to DIN-EN-ASME-JIS
"

Vibration resistance

-

Horn antenna

4 g at 5 … 200 Hz according to EN60068-2-6
(vibration with resonance)

-

Parabolic antenna

1 g at 5 … 200 Hz according to EN60068-2-6
(vibration with resonance)

Shock resistance

-

Horn antenna

4)

Not with steam

5)

Not with steam

VEGAPULS

62 • Approval according to LPR radio standard

65

10 Supplement

41722

-EN

-120706

Advertising