Bray 6A Series O&M Manual User Manual

Page 13

Advertising
background image

All information herein is proprietary and confidential and may not be copied or reproduced without the expressed written consent of BRAY INTERNATIONAL, Inc.

The technical data herein is for general information only. Product suitability should be based solely upon customer’s detailed knowledge and experience with their application.

Series 6A Operation & Maintenance – Calibration & Commissioning

6A O & M : 13

Automatic (AUT)

Automatic is the standard mode. In this mode, the positioner

compares the setpoint position with the actual position. The

positioner moves the actuator until the control deviation reaches

the configurable dead zone. A fault message is displayed if the

dead zone cannot be reached.
Diagnostics

Proceed as follows to call the “Diagnostics” mode from the

“Automatic” or “Manual” modes:
Press the three buttons of the positioner at the same time for at

least 2 seconds.
Current operating data can be called and displayed in this mode,

e.g.:

• Stroke number
• Number of changes in direction
• Number of fault messages

Note: The “Automatic” and “Manual” modes remain set when

switching to the “Diagnostics” mode. The positioner responds

as per the set mode:

• The predefined setpoint is used as a control variable in the

automatic mode.

• The last reached position is retained in the manual mode.

Using the built in Diagnostics
The S6A has various monitoring functions with which changes on

the actuator and valve can be detected and signaled if applicable

when a selectable limit has been exceeded. This information may

be important for diagnosis of the actuator or valve and will help

to schedule preventative maintenance. The measuring data that is

monitored is listed below:

• Travel integral
• Number of changes in direction
• Alarm counter
• Self-adjusting dead zone
• Valve end limit position (for detection of valve seat wear or

deposits)

• Operating Hours (also according to temperature and travel

ranges) as well as min./max. temperature

• Operating Cycles of piezoelectric valves
• Valve positioning time
• Actuator leakages

The diagnostics display has similar structure to that of the

“Configuration” mode:

• The upper line shows the value of the diagnosis variable.
• The lower line shows the number and the abbreviation of the

displayed variable.

Some diagnostics values can be greater than 99999. In such a

case, the display switches over to the exponential view. Example:

The value “1234567” is shown as “1.23E6”.
In the event of a fault an error code will be shown on the lower

left hand corner of the display. There is a fault table in the

“Troubleshooting” section of this manual which explains what

each of these codes represents.
The following table provides an overview of values that can be

displayed. The last column contains “X” if the value can be set

to zero.

Advertising