Max Machinery 121 INDICATOR User Manual

Page 48

Advertising
background image

Pg 48

121-000-350 © 2001 Max Machinery, Inc.

ENTER

STOP

ENTER

STOP

8.2.10
Calibrate CH1 0V
Submenu Group

8.2.11
Calibrate CH1 10V
Submenu Group

Calibrate ch1 0V
Vin=TB1-2 GND=TB1-4

Calibrate ch1 0V

0 CALIBRATING ——

Calibrate ch1 0V

*** DONE ***

Calibrate ch1 0V
Iin=TB1-2 GND=TB1-4

Calibrate ch1 10V
Iin=TB1-2 GND=TB1-4

Calibrate ch1 10V

0 CALIBRATING ——

Calibrate ch1 10V

*** DONE ***

Calibrate ch1 10V
Iin=TB1-2 GND=TB1-4

Advance to

Calibrate ch2 0V

To Calibrate: Connect Voltage Source (+) TB1-2, (-) TB1-4.
Input 0V and press Enter.

This message is displayed during calibration.

This message is displayed when the 0V calibration is fin-
ished.

The display will automatically return to the Calibrate CH1 0V
submenu. Press the Down arrow key to advance to the CH1
10V calibration.

To Calibrate: Connect Voltage Source (+) TB1-2, (-) TB1-4.
Input 10V and press Enter.

This message is displayed during calibration.

This message is displayed when the 10V calibration is fin-
ished.

The display will automatically return to the Calibrate CH1 10V
submenu. Press the Down arrow key to advance to the CH2
0V calibration.