Specifications – Max Machinery 121 INDICATOR User Manual

Page 5

Advertising
background image

121-000-350 © 2001 Max Machinery, Inc.

Pg 5

Specifications

Specifications:
Environmental

Indoor Use

Altitude up to 2000m

Operating Temperature: 0C to +50C
Storage Temperature: -40C to +85C
Maximum Relative Humidity : 80% for temperatures

up to 31C decreasing linearly to 50% RH at 40C

Main supply voltage fluctuations not to exceed ±10%
of the nominal voltage

Transient overvoltage according to INSTALLATION

CATEGORY II (see UL 3101-1 Annex J)
POLLUTION DEGREE 2 in accordance with
IEC 664 (see 3.7.3)
Materials: UL, CSA, VDE approved

Approvals: CE Approved Light Industrial,

UL File #: E192404
CSA Pending

Display

Type: 2 lines of 20 characters
Type: VFD
Character Size: 0.3" nominal

User selectable label descriptors and units of measure

Keypad

Keypad Type: Membrane Keypad
Keypad Rating: Sealed to NEMA 4
Number of keys: 16

Enclosure

Size: See Dimensions on page 9
Depth behind panel: 6.5" including mating connector

Type: DIN
Materials: Plastic, UL94V-0, Flame retardant

Bezel: Textured per matt finish
Equipment Labels: Model, safety, and user wiring

Power Input

The factory equipped power option is internally fused.
An internal line to line filter capacitor is provided for
added transient suppression.
121-200: 110VAC: 85 to 127 Vrms, 50/60 Hz

121-201: 12VDC: 10.5 to 14 VDC

Control Inputs

Switch Inputs are menu selectable for Start, Stop,

Reset, Lock, Inhibit, Alarm Acknowledge, Print or Not

U s e d .

Control Input Specifications

Input Scan Rate: 10 scans per second
Logic 1: 4 - 30 VDC
Logic 0: 0 - 0.8 VDC
Transient Suppression: 500 V fast transient

(Capacitive Clamp)

Input Impedance: 100 K
Control Activation: Positive Edge or Pos. Level
based on product definition

Flow Inputs:
Analog Input:
Ranges

Voltage: 0-10 VDC, 0-5 VDC, 1-5 VDC
Current: 4-20 mA, 0-20 mA

Basic Measurement Resolution: 16 bit
Update Rate: 5 updates/sec minimum

Automatic Fault detection: Signal over/under-range,

Current Loop Broken
Calibration: Self Calibration and Auto-zero continuously
Extended calibration: Learns Zero and Full Scale
of each range using special test mode.
Fault Protection:
Fast Transient: 500 V Protection

(Capacitive Clamp)

Reverse Polarity: No ill effects
Over-Voltage Limit: 50 VDC Over voltage protection
Over-Current Protection: Internally current limited
Protected to 24 VDC.
Compensation Input
The compensation input is menu selectable for
temperature, density or not used.
Operation: Ratiometric
Accuracy: 0.01% FS
Thermal Drift: Less than 100 ppm/

o

C

Basic Measurement Resolution: 16 bit
Update Rate: 1 update/sec minimum
Automatic Fault detection:

Signal Over-range/under-range
Current Loop Broken
RTD short
RTD open
Fault mode to user defined default settings
Transient Protection: 500 V (Capacitive Clamp)
Reverse Polarity: No ill effects
Over-Voltage Limit (Voltage Input): 50 VDC

Available Input Ranges
Voltage: 0-10 VDC, 0-5 VDC, 1-5 VDC
Current: 4-20 mA, 0-20 mA
Resistance: 100 Ohms DIN RTD
100 Ohm DIN RTD (DIN 42-760, BS 1904):
Three Wire Lead Compensation
Internal RTD linearization learns ice point
resistance
1 mA Excitation current with reverse polarity protection
Temperature Resolution: 0.01 C
Excitation Voltage
110/220 VAC Powered Units
Menu Selectable: 5, 12 or 24 VDC @ 100mA
12 VDC Powered Units
5 VDC @ 100mA