Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual

Page 108

Advertising
background image

103

4.0 Specifications


Input impedance: > 10 megohm
Warm-up time:

≤ 30 Seconds to full rated accuracy

Ohm Measurement Accuracy:
Temperature Effect:

≤ 0.001 ohm / °C

RTD Measurement Accuracy: See RIO Accuracies Table
RTD Simulation Accuracy: See RIO Accuracies Table

Resolution:

± 0.1° or ± 1° for RTDs

Resistance range: 0 – 4000 ohms
Temperature:
Storage: -40

° F to 140° F (-40° C to 60° C)

Operating: 14

° F to 122° F (-10° C to 50° C)

Connectors – Measure temperature: standard connector for reference RTD or Measure Cable for general purpose
measurements on installed RTDs
Simulate resistance: User Cable with connector and two alligator clips
Power: supplied from MFT (simulate/source mode only)
Units:

°F, °C, °R, °K, ohms

Weight: 3 ozs

5.0 Accessories for RIO4000

P/N A900028-90502

Simulation Cable for RTD 2-Wire simulation.







P/N A900028-90200

Mating plug connector for use with customer’s dedicated RTD







P/N A900028-90500

Measure cable assembly; RTD connector, 5 ft. copper cable, four (4) alligator clips







P/N A900028-90201

Pt 100 RTD, DIN Class "A", 0.00385 alpha, 4 leads, 1/4"
O.D., SS probe, 10" probe length, 3.5" handle and 5 ft coiled cable with connector for
RIO4000 temperature module









Advertising
This manual is related to the following products: