67

the Analog Trim main menu screen. Pressing Back from the Analog Trim screen will return the MFT to
the Config Menu screen.

Select End option to return field device to original output and exit loop test. When you select End, the
screen shows:










Or, press Abort soft key from loop test main screen to exit the loop test.

Select Back key to return back to device Config Menu.

BASIC INFO
This function provides basic but valuable HART device information to the user.
Use the Up or Down soft key to move the selection arrow until it points to Basic Info, then press the Select
soft key.
The following screen will appear:













All of the options listed on this screen are accessed by scrolling the cursor Up or Down to the desired
parameter and pressing Select to enter that option’s edit screen.

Non-editable screens provide information only. The Back soft key will be the only soft key shown for these
screens. For options that can be edited, press the Edit soft key to change current settings. Use the
Alphanumeric Grids options (see Alphanumeric Entry Section in this manual) and soft keys to select desired
characters. Press the Select key to enter alphanumeric information. Press Done to save the information.
The MFT will send the update to the HART device and return to the Basic Info main menu.

Tag:
Allows the user to program a tag number/name for the HART device. Up to 8 characters, including
spaces, may be used.

Date:
Allows the user to set the date.

Descriptor:
Allows the user to provide a brief custom description of the HART device. Examples: “Pipe 7E”
or “DP xmtr”. Up to 15 characters, including spaces, may be used.

Message:
Allows the user to enter a short message, similar to the Descriptor option above. Up to 20
characters, including spaces, may be used.

Model:
Provides information on the specific model of the connected HART device. This option is non-
editable.

Hart

1151: Pump 8
Returning Fld device to
original output





Inc Dec Next Done

Basic Info


1151: Pump 8

Tag

Date
Descriptor
Message
Private Label
Revisions
Construction Matls

Up Down Select Back