Rio setup, Up down select back – Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual

Page 98

Advertising
background image

93














The user will be prompted:

“Short the Sim. Cable Leads (clip together)”


Clip the alligator clips together and then Select the

OK soft key. The RIO4000 will perform a self

calibration and then the user will be prompted:

“Connect Sim. Cable Leads to the device”


Below is an example of 2, 3, and 4-Wire RTD Configurations for an RTD Sensor to a given
manufacturer’s Temperature Transmitter. There can be different configurations, however these
examples illustrate the corresponding RIO4000 connection required to perform RTD Simulation for 2, 3
and 4 Wire RTD’s.































RIO Setup

Settings & Functions
Choose New Probe
Display: Degrees or

Simulate Mode
Measure Mode

Selected Probe:
S->Pt 100 (385)


Up Down Select Back

1

2

3

4

Sensor Connections

JUMPER

Example 4-Wire RTD

Configuration.

1

2

3

4

RIO4000

Example 3-Wire RTD

Configuration.

Example 2-Wire RTD

Configuration.

1

2

3

4

1

2

3

4

1

2

3

4

RIO4000

JUMPER

1

2

3

4

RIO4000

JUMPER

RIO4000 Connections

Advertising
This manual is related to the following products: