Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual

Page 89

Advertising
background image

84

2.4.3 Select

Low pt to enter the beginning test point value.

2.4.4 Select

Upr pt to enter the ending test point value.

2.4.5 Select

Steps to enter the number of test points in the procedure including the Upr pt.

2.4.6 Select

Dwell Time to choose the duration of each simulated point or chose Manual for the

manual advance option.

2.5 Auto-Ramp x menu option allows pre-programmed ramp-type simulations to be set up and

executed automatically by the MFT / TIO. Auto-Ramps are automatically stored for future use.



.








2.5.1 Select Start menu option to begin the indicated Auto-Ramp number.
2.5.2 Select Units to change the Auto-Ramp x simulation units.
2.5.3 Select Low pt to enter the beginning test point value.
2.5.4 Select Upr pt to enter the ending test point value.
2.5.5 Select Ramp up time to enter the test ramp up time in seconds (Low pt to Upr pt)
2.5.6 Select Ramp dn time to enter the test ramp down time in seconds (Upr pt to Low pt)
2.5.7 Select Start delay to select a start delay from None to 5 minutes.

3.0 User Re-Calibrate

This menu option accesses the controls used for re-calibration of the TIO0110 module using local standards.

This calibration requires a highly accurate DMM, a stable, programmable DC voltage source and a standard type
TT thermocouple probe in a stable ice bath. Required equipment is as follows:

DMM

Fluke 8508A, HP3458A or equivalent

DC Voltage Source

EDC 52X or equivalent

Thermocouple

Omega TRCIII ice point with TRP-T thermocouple probe or equivalent


Each selection on the TC Setup screen enables a mode to measure and capture calibration data. The values are
stored in non-volatile memory in the sensor module. Refer to the MFT TIO0110 Calibration Manual for details of
the re-calibration process.














TC Setup
User Re-Calibration
Offsets
CJ gain
TC neg gain
TC pos gain
CJ Correction
DAC coarse gain
DAC fine gain
DAC offset
Full Recal

Up Down Select Back

TC Setup
Auto-Ramp x
Start
Units:

F

Low pt:

0.000 F

Upr pt: 400.000 F
Ramp up time: 30 sec
Ramp dn time: 30 sec
Start delay: 30 sec

Up Down Select Back

Advertising
This manual is related to the following products: