Connect the leads and any necessary jumpers to the transmitter RTD sensor terminals and select the

OK

soft key. The RIO Simulation screen shown below will be displayed. From this screen the user can select
simulation units, start a simple simulation, assign a step value change, or setup and execute automated
step or ramping test procedures.

















1.2.1 Start

begins an RTD (resistance) simulation. When

Start is selected with the Select soft

key, a numeric entry screen appears to facilitate the entry of the desired simulation value
(degree or ohm value as selected at

Units line). Once a value has been entered and saved,

the MFT will use the RTD type and alpha coefficient previously specified to both determine
and output the corresponding ohm value through the User Cable.

1.2.2 Units is used to select the temperature unit for the simulation.

°C, °F, °K, and °R are the

available. The user can also select ohms if desired.

1.2.3 Step

Value is a user entered value that can be used to increment or decrement the

simulation value after the simulation is started.
















1.2.4 Auto Step 1-3 menu options enable pre-programmed Simulations to be set up and executed

automatically. Auto-Steps settings are remembered and stored for future use.



.




Select

Start menu option to begin the indicated Auto Step number.

RIO Simulation
Simulating
Start
Units: C
Step Value: 10.0 C
Auto Step 1
Auto Step 2
Auto Step 3
Auto Ramp 1
Auto Ramp 2
Auto Ramp 3

Up Down Select Back

RIO Simulation
Fixed Simulation Mode
Enter Output in C

The S->Pt 100 (385)
probe has a range of
-200.0 C to 850.0 C

_ 0.0

< > Inc Done

Fixed Simulation Value Entry Screen

RIO Simulation
Simulating Probe
U->Pt 100 (385)

Output = 0.0 C

IV: 4.000 mA


Inc Dec Change Exit

Simulation with mA/V