Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual

Page 93

Advertising
background image

88

3.0 Manual Source and Simulation Operations–

3.1 Wire the test circuit as shown in Diagram 3.1 and 3.2 depending on the desired application:



3.2 From MFT

Measure Mode, press the More soft key


3.3 Press the

Sensr soft key and then Next until the VMA module information is displayed (MFT bay S1

module information is displayed first, then Next changes to bay S2, etc…)


3.4 Press the

Setup soft key option for the VMA

module to display the VMA Main Setup Display












VMA Main Setup Display

Vma Setup
Current Settings:
Start
Mode

mA Source

Auto-Step 1
Auto-Step 2
Auto-Step 3
Step Value 1.000 mA



Up Down Select Back

VOLTS

COM

mA

Power
Supply

Loop Load

VMA

Module

-

+

+

-

Milliamp
Receiver

Diagram 3.2 – Transmitter Simulation Connection Diagram

VOLTS

COM

mA

Milliamp

Receiver

Loop Load

VMA

Module

Diagram 3.1 – mA Source & V Source Connection