Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual

Page 85

Advertising
background image

80

Sensor Modules List




For a precise list of Intrinsic Safety certified modules, please refer to the Intrinsic Safety
Control Drawing in the Appendix of this manual.










FDN0010

1 - 10" H20 dp

± 0.1% FS from 23° F to 122° F (± 0.009 H20 dp)

FDN0020

0 - 20" H20 dp

± 0.1% FS from 23° F to 122° F (± 0.020 H20 dp)

DDN0100

0 - 100" H20 dp

± 0.025 " H20 dp

DDN0200

0 - 200" H20 dp

± 0.05 " H20 dp

DDN0400

0 - 400" H20 dp

± 0.10 " H20 dp

DDN2000

0 - 2000" H20 dp

± 0.50 " H20 dp

DGI0020

0 - 20 PSIG

± 0.005 " PSIG

DGI0200

0 - 200 PSIG

± 0.05 " PSIG

DGI0500

0 - 500 PSIG

± 0.125 " PSIG

DGI1000

0 - 1000 PSIG

± 0.25 " PSIG

DGI1500

0 - 1500 PSIG

± 0.375 " PSIG

DGI2000

0 - 2000 PSIG

± 0.500 " PSIG

DAI0900

0 - 900mm Hg Abs

± 0.225 mm Hg Abs

DAI2000

0 - 2000mm Hg Abs

± 0.50 mm Hg Abs

DAI5200

0 - 5200mm Hg Abs

± 13.00 mm Hg Abs

GGI3000

0 - 3000 PSIG

± 0.05% FS from 23° F to 122° F (± 1.50 PSIG)

TIO0110

Starter Kit for TIO0110 module t/c meas. & sim. - Universal

T/C wire Kit 1 for E, J, K, N, T; mini-t/c conn., 1 m cable
t/c wire Kit 1 for B, K, R/S; mini-t/c conn., 1 m cable
mV meas. / sim. cable assbly; mini-t/c conn., 1 m cable

VMA0055

Starter Kit for VMA0055 module V /

I

meas. & sim. - Universal

RIO4000

Starter Kit for RIO4000 module RTD meas. & sim. - Universal

Measure cable; module connector by four (4) alligator clips
RTD mating connector (for use with customer's RTD)
Simulate Cable; module connector by two (2) aligator clips

V/I: Measure ±55 Vdc, Source 0-28Vdc; mA: Measure ±100 mA DC,
Source 0-100 mA dc (Recommend VMA Test Lead Kit (P/N

VMA test lead kit, banana plugs on 9" breakouts both ends, assorted
connectors

RTD measure and simulate module supports 22 different RTDs

V/mA Meas./Sim.

RTD Meas./Sim.

PRESSURE SENSOR MODULE MODEL NUMBER, RANGE & ACCURACY

Modules are Intrinsically Safe when used in accordance with Control Doc.

"D" Series Accuracy is: ± 0.025% FS from 23° F to 122° F

T/C measure and simulate module supports types
B,C,E,J,K,N,R,S,T,M,P

Measure and Simulation Module Model Number (

Modules are Intrinsically Safe when used in accordance with

Control Doc).

T/C Meas./Sim.

Advertising
This manual is related to the following products: