Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual

Page 94

Advertising
background image

89


3.5 Select the

Mode option to select from Xmitter Sim, mA Source or V Source options. Use the soft

keys to move cursor

Up or Down, to Select the desired option and to go Back to the VMA Main Setup

Display.


3.6 From the VMA Main Setup Screen, select

Start


3.7 The user will be prompted to enter the starting value with the following screen. Use the soft keys to

enter the desired value.











Inc – Increases the value shown above the cursor
Dec – Decreases the value shown above the cursor
Next – moves cursor to the right; eventually wraps back to left position
Done – Ends data entry session and proceeds with the selected operation

3.8 Once the value is entered, press the

Done soft key to initiate the selected operation.


The operation depicted below is for mA Source. Use the soft keys to change the output value as desired.
Note that the MFT base unit’s IV (current/voltage) meter reading is shown on this display for use in
applications requiring both sourcing and measurement. The IV line can be changed at this display from V
to mA to off by using the MFTs “Function” and “Units mA / V” keys as usual (see the MFT User’s Manual
section on Changing Units for more information).










Inc – Increases the current value by the Step Value (entered in the VMA Main Setup Display).
Dec – Decreases the current value by the Step Value (entered in the VMA Main Setup Display).
Chng – Allows the user to enter another value.
Exit – Ends simulation mode, returns to the VMA Main Setup Display.


4.0 Automated Operation – Source and Simulation

The VMA has features that allow step functions that can be utilized in the source/simulation options. The
standard steps programmed by the user are stored for future recall and use.

4.1 Refer to Diagrams 3.1 and 3.2 and make connections for the desired application.

4.2 From the VMA Main Setup Menu, arrow down and select

Auto-Step X.




Vma Settings
Current Settings
Enter Output Value


_ 0.000

Inc Dec Next Done

mA Source

Output = 16.000 mA

IV = 16.015 mA


Inc Dec Chng Exit

Advertising
This manual is related to the following products: