Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual

Page 81

Advertising
background image

76

APPENDIX

PRODUCT SPECIFICATIONS OVERVIEW


BASE UNIT:
9" L x 4.7" W x 3.2"D ABS plastic case with shock absorbing boot, 28 ozs. Three bays for measurement
modules

DISPLAY: 128 x 128 pixel graphic display, 2.6” x 2.6” viewable area, 5-digit resolution

POWER: 6 AA alkaline batteries (1.5 V), 6 AA NiMH batteries (1.2 V) or optional 100 – 240 Vac power adapter (P/N
9B000007)

TEMPERATURE SPECIFICATIONS:
Storage: -40° F to 140° F (-40° C to 60° C)
Operating: 23° F to 122° F (-5° C to 50° C)

HUMIDITY SPECIFICATIONS:
10- 90 % RH @ 50°C

CONNECTIONS:

Pressure: 1/8” FNPT, 316ss

Electrical: Standard banana jacks (¾” center)

HART & Loop: Standard banana jacks (¾” center)

Data transfer: DB-9


MODULES – select from a variety of positive-lock measurement modules and special function modules for excellent
versatility. Modules can be installed in any available bay. Modules should not be removed or installed while the base
unit is On. See the Intrinsic Safety Control Document shipped with each handheld for IS restrictions.

AVAILABLE UNITS –Pressure: PSI, inches H

2

O and cm H

2

O (inches H

2

O with 4° C, 20° C, and 60° F ref. temp.),

kg/cm

2

, kPa, mBars, Bars, mm Hg, inches Hg (inches Hg with 0° C ref. temp), oz/in

2

. Temperature:

°F, °C, °R, °K.

Others based on module option.

CONFIGURABLE DISPLAY – user configured for one to four measurements.

AUTO SHUT-OFF - occurs after user-selected period of keypad inactivity or can be disabled

DAMPING – user selected options include no damping, Smart Damping, and exponential damping (user selectable
time constant from 0.5 to 30 seconds to damp pulsation and stabilize the display).

DIAGNOSTICS
– MFT 4000 self-tests upon power up and upon installation of sensor modules. Advisories are posted
on the display.

NIST TRACEABLE ACCURACY:
Pressure: ± 0.025% to ± 0.10% of F.S including all effects of linearity, repeatability, and hysteresis and temperature
over the range 23° F to 122° F.
Current / Voltage: ± 0.025% of F.S. ± 0.001% of F.S. per ° F (reference temperature is 70° F)

APPROVALS: Optional Intrinsic Safety for Class I, Div I, Groups A, B, C, D certified by MET Laboratories, Inc per
UL3111-1, CAN/CSA-22.2 No. 157, CAN/CSA-22.2 No. 1010.1-92 and UL913, Fifth Edition, Rev. 2/21/97

PRESSURE MODULE LIMITS / COMPATIBLE MEDIAS:
xGI / xAI modules: 2X F.S. range / Medias compatible with 316ss
xDN modules: 2X F.S. range when pressurized on high side only,

150 PSI when applied simultaneously to high and low
sides / Medias are clean, dry gases only.




Advertising
This manual is related to the following products: