Meriam MFT 4010 Modular Calibrator_HART Communicator User Manual

Page 53

Advertising
background image

48

Select Begin Calibration once satisfied. The following screens appears:
















Page 1

Page 2 (<More>)


Use the MFT user interface to enter the Tag, S/N, and other information needed. Be sure the
Meas/Sim line is directed to the RIO4000 module location (S1, S2 or S3). Use Temperature Setup

on the bottom of the second page to setup or change the RTD probe information (units, number of
wires, RTD material, alpha coefficient, etc…). When satisfied, move the cursor arrow to the Begin
Calibration
menu and press the Select soft key to continue. The following screen will appear.

















Select the number of set points for the switch device of interest and press the Select soft key to
continue. The following screens will be displayed. Use this MFT user interface to enter the rest of the
switch test detail. Be sure to consult the information on the <More> menu since it allows the user to
specify a LRV and URV for the test. This is especially useful for switch tests using a simulation
module since these values will provide start and stop values for the simulation module.














Cal
Non-Hart Cal
Begin

Calibration

Tag:

DEFAULT


S/N:

00000

Meas/Sim: S2
LRV : 0.000
URV : 10.000
Units : C
Xfer Fn : Linear
<More>

Up Down Select Back

Cal
Non-Hart Cal
Begin

Calibration

Measure: IV
Output LRV: 19.900
Output URV: 20.000
Units:

V

Temperature Setup





Up Down Select Back

Cal
Number of Set Points:

1

2
3
4
5
6

Choose the number
of Set / Reset PAIRS

Up Down Select Back

Switch Test
Setup: As Found

Start 1 of 1

Set Pt : 7.000
Reset Pt: 6.5.000
Direction : Inc
Test Reset Pt : Yes
Ramp Minutes: 5
Start Delay : 0
Contact Form : NO
VMA (wetting) : On
<More>
Up Down Select Back

Page 1

Page 2 (<More>)

Switch Test
Setup: As Found

Start 1 of 1

LRV : 0.000
URV : 10.000







Up Down Select Back

Advertising
This manual is related to the following products: