Maintenance tests, X series overview, Chapter 1 – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 11: Chapter 1 maintenance tests

Advertising
background image

9650-001356-01 Rev. D

ZOLL X Series Service Manual

5

Chapter 1

Maintenance Tests

X Series Overview

This chapter includes step-by-step procedures for various maintenance tests. At a minimum, a
trained biomedical technician must perform a CO

2

calibration and an NIBP calibration check

once per year. A CO

2

calibration check may be performed at any time to verify the CO

2

calibration. It is necessary to perform maintenance tests after a module is repaired to ensure that
the functions of the X Series unit work properly. Use the checklist at the back of this document
(ZOLL X Series Maintenance Tests Checklist) to record your results of the maintenance tests.

Because the X Series units must be maintained ready for immediate use, it is important for
users to conduct the Operator’s Shift Checklist procedure at the beginning of every shift. This
procedure can be completed in a few minutes and requires no additional test equipment. (See
the ZOLL X Series Operator’s Guide for the Operator’s Shift Checklist.)

Advertising