Removing the EtCO2 and NIBP Modules
9650-001356-01 Rev. D
X Series Service Manual
79
8
Disconnect the 5 NIBP tubes.
9
Disconnect the 2 red
connectors, then remove the
4 screws (top 2 screws have
washers) securing the NIBP
MOD G board. Lift NIBP
MOD G board and pump out.
Note: During reassembly,
align connector on bottom of
NIBP board to monitor board.
Torque screws to 6 in-lbs.
Note: During reassembly, put
the hose on left side of fan
through the middle slot, and
the hose on the right through
in the bottom slot.