4.0 Heart Rate Display Test

Equipment

Fluke Impulse 7000DP Analyzer

ECG Cable (3-lead, 5-lead or 12-lead)

Test Setup

Press Lead quick access key until II displays.

Connect the ECG leads to the ECG studs on Analyzer.

Do this...

Observe this...

Pass/Fail

4.1 Set the ECG Simulator to 120BPM.

The Heart Rate displays as 120 +/- 4 bpm.

o

o