4.0 Heart Rate Display Test
Equipment
Fluke Impulse 7000DP Analyzer
ECG Cable (3-lead, 5-lead or 12-lead)
Test Setup
Press Lead quick access key until II displays.
Connect the ECG leads to the ECG studs on Analyzer.
Do this...
Observe this...
Pass/Fail
4.1 Set the ECG Simulator to 120BPM.
The Heart Rate displays as 120 +/- 4 bpm.