Removing the printer module – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 70
Advertising
C
HAPTER
3
D
ISASSEMBLY
P
ROCEDURES
64
9650-001356-01 Rev. D
Removing the Printer Module
Tools Required
•
#1 Phillips screwdriver with torque capability
Setup
•
This procedure should be conducted on a static mat.
•
Wear a wrist strap to ground during the procedure.
To reinstall the printer module, reverse the steps.
Printer Disassembly
1
Place the unit on the mat in a
horizontal position.
2
Open the printer housing and
remove the printer paper.
3
Place the unit screen down
and remove the three (3)
screws outside the printer
module in the illustrated
areas.
Note: During reassembly,
torque screws to 6 in-lbs.
Advertising