0 leakage current test, Leakage current test – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 27
Advertising
Maintenance Tests
9650-001356-01 Rev. D
ZOLL X Series Service Manual
21
7.0 Leakage Current Test
Equipment
See the manufacturer’s instructions or supplied specifications for the leakage tester you use.
Setup
See the manufacturer’s instructions or supplied specifications for the leakage tester you use.
Repeat leakage test with accessories: MFC, external paddles, internal paddles, and anterior/
posterior paddles. Perform these tests at the line-power voltage and frequency used in your
installation.
Maximum Leakage Acceptance Limits
Patient Leakage
Mains on Applied Part
Normal
Single Fault
Single Fault
ECG
10
50
100
MFC
10
50
100
Normal
Single Fault
Earth Leakage
2500
5000
Advertising