5.0 Heart Rate Alarm Test

Equipment

Fluke Impulse 7000DP Analyzer

Do this...

Observe this...

Pass/Fail

5.1

Press the Leads quick access key

(

)

on the X Series unit and

select lead II.

II displays next to the Leads quick
access key.

o

o

5.2

On the X Series unit:

Press the More quick

access key (

) key to

access the second set of
quick access keys.

Press the Alarm quick

access key (

).

Press the Limits softkey.

The Alarm Settings menu displays.

o

o

5.3

On the X Series unit, select HR/PR
Alarm in the Alarm Settings menu
and use the navigation keys to do
the following:

Set the Upper Limit Enable

to ON.

Set the Lower Limit

Enable to ON.

Set the Lower limit to 35.

Set the Upper limit to 150.

Use the navigation keys to exit the
Alarm Settings menu.

5.4

On the X Series unit, press the
Alarm Cancel quick access key

(

).

The red alarm suspension timer
displays at the top of the screen next
to the message area.

5.5

Remove a lead wire from the

ECG

simulator

.

T

he ECG LEAD FAULT message

displays.

o

o

5.6

On the X Series unit, press

.

The Alarm Suspend key flashes

.

5.7

Reattach ECG Lead wire to

ECG

simulator

.

The ECG LEAD FAULT message no
longer displays.

o

o

5.8

Set simulator to 30 BPM or lower.

Heart Rate Value is highlighted and
the alarm tone sounds. The HR/PR
LOW ALARM
message displays.

5.9

On the X Series unit, press

.

The Alarm Suspend key flashes.

The

red alarm suspension timer displays
at the top of the screen next to the
message area.

o

o

5.10 Set the simulator to 120 BPM.

5.11 On the X Series unit, press

.

Alarm is enabled. The alarm suspend
key flashes. The alarm suspension
timer disappears.

o

o