0 inflation test, Inflation test – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 48

Advertising
background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

42

9650-001356-01 Rev. D

23.0 Inflation Test

Equipment

Dual lumen NIBP hose, stop watch, NIBP simulator

Do This...

Observe This...

Pass/Fail

23.1 Verify that the pressure reported by the

NIBP Test control panel or the external
pressure standard system is less than 5
mmHg.

23.2 From the NIBP Test panel, use the

navigation keys to select Close Valve.

23.3 Select Turn Pump On and immediately

start the timer.

23.4 Stop the timer as soon as pressure exceeds

210 mmHg.

Verify the time is less than 7
seconds.

o

o

23.5 Select Turn Pump Off, then Open Valve.

23.6 Exit the Service Menu and turn off the

unit.

Advertising