Removing the Handle
9650-001356-01 Rev. D
X Series Service Manual
63
Removing the Handle
Tools Required
•
#1 Phillips screwdriver with torque capability
•
Wooden stick
Setup
•
This procedure should be conducted on a static mat.
•
Wear a wrist strap to ground during the procedure.
To reinstall the handle assembly, reverse the steps.
Handle Disassembly
1
Using a wooden stick, lift the
screw covers off the handle in
two locations.
2
Remove four (4) screws (2 on
each handle) and remove the
handle.
Note: During reassembly,
torque screws to 8 in-lbs.