0 etco2 test, Etco2 test, 0 etco – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 41: Test

Advertising
background image

Maintenance Tests

9650-001356-01 Rev. D

ZOLL X Series Service Manual

35

18.0 EtCO

2

Test

Equipment

CO

2

sampling line airway adapter

Do this...

Observe this...

Pass/Fail

18.1 On the X Series unit, install the airway

adapter set and sampling line in the
connector at the side of the unit.

18.2

Press the

CO

2

quick access key

(

)

to

activate the CO

2

trace.

The CO

2

trace is on the display.

o

o

18.3 On X Series unit, gently blow into the

airway adapter set while observing the
display.

The CO

2

waveform changes on the

display.

o

o

CO

2

CO

2