ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 87
Advertising
Removing the Monitor Board and SpO2 Modules
9650-001356-01 Rev. D
X Series Service Manual
81
2
Disconnect the SpO
2
ground
cable.
3
Remove 4 screws on the left
side in the illustrated areas.
Note: During reassembly,
torque screws to 6 in-lbs.
4
Turn the unit over.
5
Remove 3 screws on the left
side of the unit in the
illustrated areas.
6
Turn the unit over.