ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 26
Advertising
C
HAPTER
1
M
AINTENANCE
T
ESTS
20
9650-001356-01 Rev. D
6.5
On the X Series unit, do the
following:
•
Press the More (
) then the
Setup (
) quick access keys.
•
Use the navigation keys to
select Supervisor>Display/
Configuration>Device Info
and select Restore All Factory
Defaults.
Note: You need a supervisor
passcode to enter the Supervisor
menu. Also the unit will reboot
during the of restoring factory
defaults.
6.6 Press
the
I Measure switch on the
Battery Current Test Fixture.
Observe the current measurement on
the DMM, and verify that the reading
is between 0.47A-1.25A.
o
o
Do this...
Observe this...
Pass/Fail