Removing the Defibrillator Boards and Capacitor Bank Modules

9650-001356-01 Rev. D

X Series Service Manual

89

3

Disconnect the defib-system
flex cable from the processor
board by pulling straight up
from board.

4

Remove screw in the middle
of the plastic isolator (located
on top of the processor
board) in the area illustrated.

Note: During reassembly,
torque screw to 6 in-lbs.

5

Remove the plastic isolator
shield.