Removing the Defibrillator Boards and Capacitor Bank Modules
9650-001356-01 Rev. D
X Series Service Manual
89
3
Disconnect the defib-system
flex cable from the processor
board by pulling straight up
from board.
4
Remove screw in the middle
of the plastic isolator (located
on top of the processor
board) in the area illustrated.
Note: During reassembly,
torque screw to 6 in-lbs.
5
Remove the plastic isolator
shield.