Removing the etco2 and nibp modules, Removing the etco, And nibp modules – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 82

Advertising
background image

C

HAPTER

3

D

ISASSEMBLY

P

ROCEDURES

76

9650-001356-01 Rev. D

Removing the EtCO

2

and NIBP Modules

Tool Required

# 1 Phillips screwdriver with torque capability

Wooden stick

Large diagonal cutters

Metal Pick

Setup

This procedure should be conducted on a static mat.

Wear a wrist strap to ground during the procedure.

Complete these procedures in the following order before performing this procedure:

1.

“Removing the Handle” on page 63

2.

“Removing the Printer Module” on page 64

3.

“Disassembling the Main Unit” on page 66

4.

“Removing the Front Enclosure Module” on page 70

To assemble the EtCO

2

and/or NIBP modules, reverse the steps.

EtCO

2

and/or NIBP Disassembly

1

Unplug the power connector
by press on the tab and
puling straight out.

Advertising