3. Display Description and Operation Procedure

17

[CONTINUOUS MODE]

Start/Stop measuring manually.
Example: If the setting is as follows in “CONFIG” menu (Refer to 3.7.2 CONTINU):

INTERVAL :

N/A

SAMPLE T.

: N/A

CYCLES

: N/A

A continuous measurement is performed until you manually stop.

Measurement Flow








[INTERVAL MODE]

Configure INTERVAL, SAMPLE T and CYCLES, and repeat measuring.
Example: If the setting is as follows in “CONFIG” menu (Refer to 3.7.3 INTERVAL):

INTERVAL :

00:01:00

SAMPLE T.

: 00:00:15

CYCLES

: 20

A total of 20 samples are performed at one minute intervals.
Note: Sampling time is 15 seconds and standby time is 45 seconds.


Measurement Flow











If the time period between two samplings is 10 seconds or longer, the pump will be turned OFF
after a sampling is finished. Then the pump will be turned ON again 10 seconds before the next
sampling starts.


Sampling Time

Start

Measurement

Finish

Measurement

Wait Time

STOP

Interval

Sampling Time
(15sec)

The 1

st

Measurement

The 2

nd

Measurement

The 20

th

Measurement

Total Time: 19min 15sec

Start

Measurement

Finish

Measurement

Wait Time

Interval

Sampling Time
(15sec)

Sampling Time

(15sec)