KANOMAX 3900 Portable Particle Counter User Manual

Page 64

Advertising
background image

3. Display Description and Operation Procedure

57

· ALARM SETTING

In ALARM SETTING set threshold for particle
alarm and option sensor alarm.

In the PARTICLE tab, configure the unit and
threshold for each particle.
Check the particle size that you want to alarm by
external contact output and audible alarm*.
(* Note that audible alarm will be activated only
when “Alarm Sound” is checked.)
Also enter the threshold that you want to alarm.
To enter the threshold tap entry field to display a
numeric keypad. Enter a value and tap [Enter].

In the OPTION tab, you can set alarm threshold for
TEMPERATURE, HUMIDITY, VELOCITY and
DIFFERENTIAL PRESSURE.

Tap the entry field to display a numeric keypad.
Enter value and tap [Enter].

To activate audible alarm check “Alarm Sound”.
Tap [APPLY] to save the setting value that is
displayed at the moment (Authentication required).

Tap [OK] to save the setting values both in
PARTICLE tab and OPTION tab, and to close the
window (Authentication required).

Tap [CANCEL] to discard the configured settings,
and the previous settings will be applied.