KYORITSU 5000 Instruction Manual User Manual

Page 22

Advertising
background image

― 20 ―

4. Pre-set current value at channel 1 is

displayed.
Can be set at every 1mA from 0 to 1000mA

¡

Press the

button and proceed to the

next step when not changing the setting.

¡

To change the setting, press the

button.

Then the indication on the LCD flashes.
Press the

or

button to set the

value to the desired one. Then press the

button to confirm it.

Note

After 1 data is recorded, current will not be

detected until current drops to 50% or less of
the preset current value as the last detected
leakage current is remaining.
Please set it to the appropriate value with
reference to the measurement performed
before starting a recording.

5. Confirm or change the preset current value on

channel 2 and 3 as well.

6. State of One-time system is displayed.

on: Recording stops when memory becomes

full.

off: Overwrite the old data, and store the latest

data.

¡

Press the

button and proceed to the

next step when not changing the setting.

¡

To change the setting, press the

button.

Then the indication on the LCD flashes. Press
the

or

button to set the value to

the desired one. Then press the

button

to confirm it.

7. Now Setting 1 is complete; "End" is displayed

on the LCD. Press the

button to return

to the screen on which "SEt.1" is displayed.

8. Press the

button to get the instrument

ready for a measurement.