― 38 ―

10. Specification

¡Measuring Ranges and Accuracy

* Continuous recording mode [RMS] (50/60Hz,Sine wave)

Crest Factor ≦ 2.5 : Sine wave accuracy +2%+5dgt

* Event/Max. value/ Capture recording mode [RMS] (50/60Hz,Sine wave)

* Current detection (Event/ Max. value/ Capture recording mode):
* Capture recording mode [instantaneous value]:

NOTE: Electromagnetic compatibility(EMC)

EN61000-4-2 Electrostatic discharge immunity(ESD)

Performance criteria: B

¡

Operating system

:

Successive Approximation

¡

Input

:

AC voltage(AC100mV/A)

¡

Rated max. working voltage : AC170mVrms, 250mV peak value

¡

Number of input channel

:

3

¡

Measuring method

:

True RMS

¡

RMS Measuring interval

:

Continuous recording mode:

approx. 1sec. to 60 min. depends on recording Interval.
(intermittent sampling)

Max. value, Event mode

:

approx. 100ms. Normally, sampling at 3.3ms intervals. (Current
detection: at the interval of about 1.6ms.)

Capture recording mode

:

approx. 100ms. Consistently , sampling at 1ms intervals.
(Current detection: at the interval of about 0.5ms.)

Range

100mA

1000mA

0∼100.0mA

0∼1000mA

Measuring

range

Accuracy

of the instrument

Accuracy when

combining with sensor

±1.0%rdg±5dgt

±2.0%rdg±10dgt

±2.0%rdg±6dgt

Range

100mA

1000mA

0∼100.0mA

0∼1000mA

±1.5%rdg±7dgt

±2.5%rdg±12dgt

±2.5%rdg±8dgt

Range

100mA

1000mA

0∼100.0mA

0∼1000mA

±3%rdg±2%fs

±4%rdg±2.5%fs

±4%rdg±2%fs

Measuring

range

Measuring

range

Accuracy

of the instrument

Accuracy

of the instrument

Accuracy when

combining with sensor

Accuracy when

combining with sensor