10.4.2 Setting the output range

(1) Press the

or

key on the RANGE screen and

select “OUTPUT RANGE” from “OUTPUT”.
Press the

ENT

key, and the OUTPUT RANGE screen

is displayed

(1) Select “RANGE” on the OUTPUT RANGE screen

and press the

ENT

key and the cursor moves to

“KIND”.

(2) Set the kind of output range (velocity, fl ow rate and

thermal fl ow).
Press the

ENT

key, and the screen appears, enabling

you to select the output range.
Select the kind of output range by the

or

key

and press the

ENT

key.

For the fl ow rate of the output range kind, set 10.3.3
of "Calorie mode" to "Used"

100

%

Flow rate

Forward

Reverse

Analog output

20mA

н100%

н20%

4mA

Output limit high

Output limit high

Output limit low

0

100%

200mA

4mA

Output limit low

Flow rate

Single range: Output one way by 0 to 100%

Output = Momentum flow rate ×100

/ Full scale

Bi-directional range: Also output back-flow by 0 to 100%.

When switching flow direction, Hysteresis
will be 10% in full scale.

(1) “RANGE”: when setting kind of output range, range type, full scale value

and output limit value.

Bulletin F-9-PUX2