7.10 Selection of sensor mounting method

Mounting methods available for the sensor are
V method and Z method as illustrated.
To select the mounting method;
Press the

key, so the “SENSOR MOUNT” is

reversed from white to blue.
Press the

ENT

key. The “SENSOR MOUNT”

screen will appear.

Select either V or Z method by the

or

key.

Remarks

Select the V method generally. Use the Z method in

the following cases:

• Ample space is not provided.

• High

turbidity

• Weak receiving waveform

• Thick scale is deposited on the pipe internal surface.

V method

Z method

(Except.

Small sensor PSX2-B)