8.2

Selection of detector

(1) Selection of mounting methods

There are 2 methods for mounting the detector; V method and Z method. For the mounting
space, see the following sketch.

Length of frame of each sensor

Frame

Length of frame of each sensor

L

L

L: Mounting dimension

<Small diameter sensor, small sensor or high-temperature sensor>

mounting dimension for
sensor displayed on the
SITE SETUP screen

D

D

<Large/Medium sensor>

Detector

L

V method

Detector

L

Z method

V method

Z method

Mounting method

Employ the Z method in the following cases.
• Mounting space need be saved (mounting space of the Z method is about one half of the V

method's).

• Turbid fl uid such as sewage is to be measured.
• Pipe has mortar lining.
• A thick fi lm of scale may have been formed on the inner surface of pipe because it is old.
• In sufi cient received signal-strength with mounting detectors in V method while using maxi-

mum transmission voltage.

Bulletin F-9-PUX2