10.1.2 ZERO ADJUSTMENT: when performing zero adjustment

[Operation]

(1) Select “ZERO ADJUSTMENT” by the

or

key

and press the

ENT

key. The zero adjustment screen

will appear.

(2) Select ZERO ADJUSTMENT, and press the

ENT

key.

Zero adjustment to be specifi ed is carried out.

• [SET ZERO]

Perform zero adjustment in situation where the fl ow
is stopped.
The measurement indication should be at zero when
the

ENT

key is pressed.

This zero calibration operation should be performed
after stopping fl ow.

• [CLEAR]

Adjustment is cleared.

On this screen, zero point is set or cleared.

When PROCESS SETTING or measurement
method (page 93) is changed, perform zero
adjustment.