• Connect the Coupler, Low PIM Load and Power Meter to the PIM 31 test

system (use torque wrench)

• Set the PIM 31 to Analyzer Mode
• Set frequencies of carrier signals 1 and 2 to the start and end frequency of

the provided frequency band.

• Set the power level of each signal to 40 dBm (20W) – with ALC: ON

Switch RF Power ON

• With offset compensation of the coupler (30dB) the power meter should

read 43dBm +/- 1.0dB

If the reading deviates, switch RF power OFF, disconnect connections,

reconnect and repeat the test.

.


If the reading is off by more than 1 dBm, the deviation could be caused
by an malfunctioning attenuator or amplifier in the PIM 31 Test System.
In this case, contact your nearest Boonton Service Center

.



With the reading within limits, continue procedure:

• Increase Signal Levels by 1dB,
• Measure power
• Repeat until Signal Levels are both at 43dBm.
• With offset compensation of the coupler (30dB) the

power meter should

read now: 46dBm +/- 1.0dB

Switch “RF Power OFF