• Read RSSI on PIM31; value should be +/-1 dB of to Signal Generator’s

Power level

• Repeat the process you are no longer able to decrease the signal

frequency F2

.

• Set PIM 31 Signal frequencies F1 and F2 back to Start and End

frequencies of PIM31 band.

• Set Signal Generator to Frequency that results for IM3 product of

frequencies F1 and F2

• Read the RSSI value on PIM31; the value should be equal to +/-1 dB of

the Signal Generator’s Power level

.

• Change the power level of the Signal Generator by 5dB.

Never exceed -60dBm

• Read RSSI on PIM31; value should be equal to +/-1 dB of the Signal

Generator’s Power level

.

• Repeat until power level range of -60dBm to -120dBm has been covered
• Switch RF Signal of Signal Generator OFF

If the readings from the previously listed procedures deviate by more than
+/- 1 dB, contact your nearest Boonton Service Center

.