D.2 command reference – BNC 1105 Universal Counter User Manual

Page 106

Advertising
background image

106

106

106

106

DISPlay:TEXT “CALC2”


D.2 Command Reference

:ABORt

Aborts any measurement in progress as quickly as possible.

Notes

When this command is sent while the measurement cycle is idle

(:INIT:CONT OFF and pending operation flag is false), it is ignored.

When this command is sent while a single measurement is in progress

(:TRIG:COUN:AUTO OFF or :CALC3:AVER OFF, :INIT:CONT OFF, and

pending operation flag is true), the measurement will be aborted and

pending operation flag set false.

When this command is sent while repetitive measurement are being taken

(:INIT:CONT ON), the current measurement in progress will be aborted and

the pending operation flag set false. A new measurement will then be

automatically initiated and the pending operation flag set true.

When this command is sent while a block of measurements is in progress

(:TRIG:COUN:AUTO ON and :CALC3:AVER ON, :INIT:CONT OFF, and the

pending operation flag is true), the measurement block will be aborted, the

pending operation flag set false and the Measuring bit in the Operation

Status Register will be set false.

Terminating a measurement in progress invalidates the result, but is the

appropriate way to terminate an automatically armed Totalized

measurement.

The command :CALCulate[1]

Performs data processing or transfer on data of scale and offset
operation acquired by the :SENSe command. Be sure to
send :CALCulate[1]:MATH:STATAe ON to activate the operation before
any :CALCulate[1] command.

:CALCulate[1]:DATA?

Queries the measurement result using scale and offset calculation.

Return

Results are formatted according to :FORMat[:DATA] ASCii|REAL setting.