BNC 1105 Universal Counter User Manual

Page 128

Advertising
background image

128

128

128

128

Sets or queries the state of the low-pass filter.

Return

A single ASCII-encoded byte. 0 means OFF and 1 means ON.

Note

This will be set to OFF after *RST command

:INPut[1|2]:FILTer[:LPASs]:FREQuency?

Queries the cut-off frequency of the low-pass filter.

Return

Numeric data transferred as ASCII bytes in <NR3> format with six

significant digits.

A value of 100E+3 is returned.

:INPut[1|2]:IMPedance <numeric value> [OHM]

Sets or queries the input impedance. It is either 50Ω or 1MΩ.

Return

Numeric data transferred as ASCII bytes in <NR3> format with six

significant digits.

Note

The input impedance will be set to 1E6 OHM after *RST command.

:INPut3:COUPling?

Queries the channel 3 input coupling.

Return

A sequence of ASCII-encoded bytes:AC.

:INPut3:IMPedance?

Queries the channel 3 input impedance.

Return

Advertising