BNC 1105 Universal Counter User Manual

Page 148

Advertising
background image

148

148

148

148

Internal 1 is used.

This command does not apply on Rise/Fall time, Duty cycle or Pulse width

measurements.

[:SENSe]:TINTerval:ARM:ESTARt[:LAYer[1]]:SOURce IMMediate

︱ TIMer ︱

INTernal[1]

Sets or queries the start arm delay fro Time Interval measurements.

Return

A sequence of ASCII-encoded bytes. IMM means Immediate, TIM means Timer and

INT means Internal.

Note

This will be set to Immediate after the counter has been turned off or a remote

interface reset.

This command does not apply on Rise/Fall time, Pulse width or Duty Cycle

measurements.

[:SENSe]:TINTerval:ARM:ESTART[:Layer[1]]:TIMer <numeric value> [S]

Sets or queries the delay time in start arm for Time Interval measurements.

Return

Numeric data transferred as ASCII bytes in <NR1> format.

Notes

This will be set to 100E-9 seconds after the counter has been turned off or a remote

interface reset.

This command only applies when [:SENSe]:TINT:ARM:ESTART[:LAYer[1]]:SOUR

TIM is used.

This command dos not apply on Rise/Fall time, Pulse width or Duty cycle

measurements.

[:SENSe]:TINTerval:ARM:ESTOP:LAYer2:SLOPe POSitive

︱NEGative

Sets or queries the slope of the external stop arm signal used in external arming Time

interval measurements.

Return

A sequence of ASCII-encoded bytes. POS means Positive edge and NEG means

Advertising