A string with the form of “<function> <parameter> [,<source list>]” is

returned.

If there is not any specification before this query, an empty string is returned

and an error -221,"Settings conflict" is generated.

The response is unaffected by *RST, recall, or [:SENS]:FUNC.

Note

Refer to (The Command :MEASure?) later in this section for the descriptions

of the <parameters> and the <source list> for each function.

Device Clear

Instructs the counter to clear the input buffer and Output Queue, resets
the parser, execution control and response formatter. Clears any
command that would prevent processing a command *RST or others.
Disables the effect of previously executed the command *OPC and
terminates the hold-off action of a *WAI, *OPC? Or data query
(:MEASure query, :READ query, :FETCh
query, :CALC:DATA?, :CALC2:LIM:FAIL?) waiting for pending operation
to complete.

Note: Diagnostics or calibration initiated from front panel may be terminated it

is waiting for user input.

:DISPlay:ENABle <Boolean>

Sets or queries if the whole display is enabled.

Return

A single ASCII-encoded byte. 0 means Disables and 1 means Enabled.

Notes

The display will be set back to Enabled after *RST command.

This value is unaffected by save&revall.

:DISPlay:MENU[:STATe] OFF

Disables the menu display and shows the results. Response indicates whether

menu or results is in display.