BNC 1105 Universal Counter User Manual
Page 131
131
131
131
131
specified value. When they are specified in voltage, auto-trigger is disabled and the
trigger level is the specified value.
<lower_reference>
default
10 pct
Percent range
0 to 100 pct
Percent resolution
10%
Voltage range
-5.125V to 5.125V for ×1 attenuation
-51.25V to 51,25V for ×10 attenuation
Voltage resolution
0.0025V for ×1 attenuation
0.025V for ×10 attenuation
<upper_reference>
default
90 pct
Percent range
0 to 100 pct
Percent resolution
10%
Voltage range
-5.125V to 5.125V for ×1 attenuation
-51.25V to 51,25V for ×10 attenuation
Voltage resolution
0.0025V for ×1 attenuation
0.025V for ×10 attenuation
<source list>
default
(@1)
range
(@1)
:MEASure[:SCALar][:VOLTage]:FREQuency?
[<expected_value> [,<resolution>]][,(@1)|(@2)|(@3)]
Sets the measurement arming to “digits” and measures the frequency.
The <expected value> and <resolution> parameters are used to configure the number of digits
of resolution arming setting.
The Channel 1 and Channel 2 share the same trigger setting so that the measurement channel
has auto-trigger enabled at 50% with a positive slope.
<expected value>
Ch1, Ch2
default
10 MHz
Range
0.100 Hz to 500 MHz
Resolution <expected value> should be within 10% of input frequency
for optimum arming configuration
<expected value>
Ch3
Default
500 MHz
Range
375 MHz to 6GHz
Resolution <expected value> should be within 10% of input frequency
for optimum arming configuration
<resolution>
Ch1, Ch2
Default
Value which indicates 4 digits of resolution for the specified
<expected value>
Range
1E-16 Hz ~ 1E6 Hz (15 to 3 digits of resolution)