The commands :measure – BNC 1105 Universal Counter User Manual

Page 129

Advertising
background image

129

129

129

129

Numeric data transferred as ASCII bytes in <NR3> format with six

significant digits.

The value returned is 50.

The Commands :MEASure?


The command :MEASure is the simplest way to make a measurement. It
configures the counter, initiates the measurement and queries the result
all in one command.

:MEASure[:SCALar]:<function>? <parameters> [,<source_list>]


The above command is the generic form of the :MEASure? Commands.
Skipping the <function> and the <parameters> will instruct the counter
to use defaults. The description of each command :MEASure? is listed
below.

:MEASure[:SCALar][:VOLTage]:DCYCle? [<reference>][,(@1)]

Sets the measurement arming to “auto” and measures Duty Cycle with the specified

parameters.

The <reference> specifies the point on the pulse where the duty cycle is determined, and

it can be specified as either a percentage or an absolute voltage. The default is in

percentage.

The <reference> parameter is used to configure the trigger setting. When it is specified

in percentage or default, auto-trigger is enabled and trigger level is the specified value.

When it is specified in voltage, auto-trigger is disabled and the trigger level is the

specified value.

<reference>

default

50 pct

Percent range

0 to 100 pct

Percent resolution

10%

Voltage range

-5.125V to 5.125V for ×1 attenuation

-51.25V to 51,25V for ×10 attenuation

Voltage resolution

0.0025V for ×1 attenuation
0.025V for ×10 attenuation

<source list>

default

(@1)

range

(@1)