toggle between positive and negative. This is to specify the slope of
the start arm signal. Press the softkey under STOP to toggle between
positive, negative and time. This is to specify how measurement
should stop. Positive and negative specifies the slope of the stop arm
signal. Time is to set the duration of measurement. As time is
selected, use the numeric keypad or the arrow keys to input a desired
duration value (refer to Ch.4.2.1). The time unit will be shown on the
top-right corner.

Remote Interface Command

Auto arming:

[:SENSe]:TOTalize:ARM[:STARt]:SOURce IMMediate

[:SENSe]:TOTalize:ARM:STOP:SOURce IMMediate

Time arming:

[:SENSe]:TOTalize:ARM[:STARt]:SOURce IMMediate

[:SENSe]:TOTalize:ARM:STOP:SOURce TIMer

[:SENSe]:TOTalize:ARM:STOP:TIMer <numeric value> [S]

External arming:

[:SENSe]:TOTalize:ARM[:STARt]:SOURce EXTernal

[:SENSe]:TOTalize:ARM[:STARt]:SLOPe POSitive|NEGative

[:SENSe]:TOTalize:ARM:STOP:SOURce EXTernal

[:SENSe]:TOTalize:ARM:STOP:SLOPe POSitive|NEGative

[:SENSe]:TOTalize:ARM:STOP:SOURce TIMer

[:SENSe]:TOTalize:ARM:STOP:TIMer <numeric value> [S]

For time interval measurement:
Press the softkey under GATE to toggle between auto and external. Press
the softkey under DONE to finish the selection.
■ If Auto Arming is selected, the counter automatically starts

continuous measurements. Besides, users can press the softkey
under Delay* to toggle between None, Time and Event. In the
settings of Time and Event, users also can input parameters

STARt:SOURce STOP:SOURce
IMMediate

IMMediate

Auto

----

IMMediate

TIMer

Time

----

EXTernal

EXTernal

EXTERNAL

POS,NEG

EXTernal

TIMer

EXTERNAL

Time

Front-Panel Gating Settings

GATE STOP