[:VOLTage]:PHASe? [(@1),(@2)]

[:VOLTage]:DCYCle? [<reference>][,(@1)]

[:VOLTage]:PDUTycycle? [<reference>][,(@1)]

[:VOLTage]:TOTalize:TIMed? [<gate_time>][,(@1)]

The CONFigure Command
The CONFigure command offers a little more flexibility than the command
MEASure? and is usually used with the commands READ?. The universal
counter sets the parameters for the requested function, but does not make
measurement. Users have an option to change the configuration. To initiate
the measurement, READ? Or INITiate commands can be used.

:CONFigure[:SCALar]:<function> <parameters> [,<source_list>]

For more information about <function>, <parameters> and <source_list>, please

refer to the above table.

If you skip the parameters or the function, or the source list, or use the

keyword DEFault instead, the default settings (parameters, function and

source) will be used.

Using this command automatically turns off the statistics operation, scale

and offset operation and limit testing.

If the trigger level is not specified in <parameters>, then auto-triggering will

be enabled and the trigger levels will be automatically set in all functions but

Voltage Peaks and Totalize on the measurement channel

.

:CONFigure

:CONFigure[:SCALar]

[:VOLTage]:FREQuency [<expected_value> [,<resolution>]]

[,(@1)|(@2)|(@3)]

[:VOLTage]:FREQuency:RATio [<expected_value> [,<resolution>]]

[,(@1),(@2)|(@3)]|[,(@2)|(@3),(@1)]

[:VOLTage]:PERiod [<expected_value> [,<resolution>]]

[,(@1)|(@2)|(@3)]

[:VOLTage]:RISE:TIME [<lower_reference> [,upper_reference]][,(@1)]

[:VOLTage]:RTIMe [<lower_reference> [,upper_reference]][,(@1)]

[:VOLTage]:FALL:TIME [<lower_reference> [,upper_reference]][,(@1)]

[:VOLTage]:FTIMe [<lower_reference> [,upper_reference]][,(@1)]