Hmg 3010 – HYDAC HMG 3010 User Manual

Page 52

Advertising
background image

HMG 3010

Page 52

Edition 11/2011 - V04 R01

Part No.: 669855

• In order to also manually generate a

measured value line, set a second event
to Actuation of button/key.

• Automatic

measurement:

Deactivated

When measurement starts, a new menu win-
dow Event Log Recording — Status opens,
enabling you to track the measurement on
the basis of the settings. Data is recorded
only when an event occurs.
The sampling rate for the event log is auto-
matically determined by the HMG in keeping
with the number of active measurement
channels:

• 0.1 ms max. 2 analogue input channels
• 0.2 ms max. 4 analogue input channels
• 0.5 ms all 10 input channels
• 1.0 ms for SMART sensors

For the above example this means that the
stamping press is switched on, the pressure
reaches 131 bar and drops back to 0 bar.
When reaching 5 bar in the falling direction,
a measured value line is generated and the
Min/Max values are reset. The last Max
value is retained in the measured value line.
A die closing pressure of 129 bar is reached
in the next cycle. This value is also written to
a new line as the pressure drops to 5 bar.

The function bar offers you the option of hav-
ing the current measured values displayed
(Measured Values) or a Table shown.

The Table option will list the data records of
each individual event
line by line. Apart
from the date and time of the respective re-
cording, the trigger event and the Min and
Max values for the active sensors are
shown. The most recent data record is
shown in the bottom line of the table.