GX-2009 Operator’s Manual

Start Up • 14

If calibration is not due or Cal. Limit Check is set to None, then the
following screen appears for a few seconds indicating when the next
calibration is due.

NOTE: The following screen only appears if Bump Test Limit Display is

set to On using the User Setup Program. The standard factory
setting for this function is Off.

4. If Bump Test Limit Display is set to On using the User Setup Program,

the next screen will depend on how Bump Limit Check is set using the
User Setup Program.

If the unit is due for bump testing and Bump Test Limit Check is set
to Confirm to use, then the following screen displays.

The alarm LED’s and buzzer will pulse several times. After this,
press and release the POWER MODE button. You will hear a beep
and startup sequence will continue.

WARNING:You must press the POWER MODE button to continue to

normal operation. If you do not press the POWER MODE
button, the buzzer will continue to beep and the LEDs will
continue to flash for 6 seconds every 5 seconds and the
unit will not enter normal operation.

If the unit is due for bump testing and Bump Test Limit Check is set
to Can’t use, then the following screen displays.

The alarm LED’s and buzzer will pulse several times. This screen
will then remain on the display until the unit is turned off. The GX-
2009 cannot be used until a bump test has been performed. See
“Performing a Bump Test in BUMP” on page 40 for bump test
procedures.

08 - 7.27

NEXT CAL

15 d

tESt

BUMP NOW

FAIL

BUMP NOW