User setup mode – RKI Instruments GX-2009 Manual User Manual

Page 52

Advertising
background image

47 • User Setup Mode

GX-2009 Operator’s Manual

User Setup Mode

This section describes the GX-2009 in User Setup Mode. User setup mode
has the same menu items as Calibration Mode with the addition of the
ALARM--P menu item which allows you to set the alarm points and the
PASSWORD menu item which allows you to turn the password feature on
or off and set the password. The common menu items are included in User
Setup Mode as a convenience to avoid turning off the unit and entering
Calibration Mode if you are updating alarm points and also want to use one
of the other menu items. In User Setup Mode, you can move through a
menu of screens to do the following:

Set the date and time (same as in Calibration Mode)

Perform a fresh air adjustment (same as in Calibration Mode)

Perform a span adjustment on all channels simultaneously (same as in
Calibration Mode)

Perform a span adjustment on one channel at a time (same as in
Calibration Mode)

Set the alarm points

Discharge the batteries with the REFRESH function (same as in
Calibration Mode)

Turn the password feature on or off and define the password

NOTE:

You can set the GX-2009 to alert you during the startup sequence
when calibration or bump testing is due with the GX-2009 Setup
Program. See the GX-2009 Setup Program Operator’s Manual for
information on setting the Cal. Limit Display and Bump Limit
Display

parameter.

The description of the menu items that are common to Calibration Mode
and User Setup Mode will refer you to the appropriate pages in the
Calibration Mode section.

Using User Setup Mode

WARNING: The GX-2009 is not in operation as a gas detector while in

User Setup Mode. Although it will respond to gas in parts
of AIR CAL, AUTO CAL, ONE CAL, and BUMP TEST, there
are no gas alarm indications.

1. Take the GX-2009 to a non-hazardous area and turn it off if it is on.