GX-2009 Operator’s Manual

Maintenance • 62

Replacing the Scrubber Filters

WARNING: Replace the scrubber filters in a non-hazardous

environment.

1. Verify that the GX-2009 is off.

2. Unscrew the two screws that secure the sensor retainer and sensor

cover retainer to the GX-2009 and remove them.

3. Using a small flat blade screwdriver, gently pry each side of the sensor

cover retainer away from the sensor retainer.

4. Remove the sensor cover and sensor retainer gasket to gain access to

the scrubbers.

5. The charcoal filter disk is located in the CO sensor diffusion port. One

filter disk is used.

One H

2

S removal filter disk is located in the combustible gas diffusion

port.

6. When the appropriate scrubber filters have been replaced, reinstall the

sensor retainer gasket and sensor cover.

POWER

MODE

Sensor Cover

Sensor Cover
Retainer

AIR

Sensor Retainer
Gasket

H

2

S

CO

%LEL

O

2

RKI

GX-2009

H2S Removal
Filter Disk

Charcoal

Filter Disk

Figure 23: Removing the Sensor Retainer and Replacing the Scrubber Filters